Abstract
A novel PET radiotracer, Flurpiridaz F 18, has undergone phase II clinical trial evaluation as a high-resolution PET cardiac perfusion imaging agent. In a subgroup of patients imaged with this agent, we assessed the feasibility and benefit of simultaneous correction of respiratory and cardiac motion. Methods: In 16 patients, PET imaging was performed on a 4-ring scanner in dual cardiac and respiratory gating mode. Four sets of data were reconstructed with high-definition reconstruction (HD•PET): ungated and 8-bin electrocardiography-gated images using 5-min acquisition, optimal respiratory gating (ORG)—as developed for oncologic imaging—using a narrow range of breathing amplitude around end-expiration level with 35% of the counts in a 7-min acquisition, and 4-bin respiration-gated and 8-bin electrocardiography-gated images (32 bins in total) using the 7-min acquisition (dual-gating, using all data). Motion-frozen (MF) registration algorithms were applied to electrocardiography-gated and dual-gated data, creating cardiac-MF and dual-MF images. We computed wall thickness, wall/cavity contrast, and contrast-to-noise ratio for standard, ORG, cardiac-MF, and dual-MF images to assess image quality. Results: The wall/cavity contrast was similar for ungated (9.3 ± 2.9) and ORG (9.5 ± 3.2) images and improved for cardiac-MF (10.8 ± 3.6) and dual-MF images (14.8 ± 8.0) (P < 0.05). The contrast-to-noise ratio was 22.2 ± 9.1 with ungated, 24.7 ± 12.2 with ORG, 35.5 ± 12.8 with cardiac-MF, and 42.1 ± 13.2 with dual-MF images (all P < 0.05). The wall thickness was significantly decreased (P < 0.05) with dual-MF (11.6 ± 1.9 mm) compared with ungated (13.9 ± 2.8 mm), ORG (13.1 ± 2.9 mm), and cardiac-MF images (12.1 ± 2.7 mm). Conclusion: Dual (respiratory/cardiac)-gated perfusion imaging with Flurpiridaz F 18 is feasible and improves image resolution, contrast, and contrast-to-noise ratio when MF registration methods are applied.
The image quality of myocardial perfusion images acquired with a PET system depends on the radiotracer used in the study. The commonly used radiotracers in cardiac PET imaging are 15O-water (H215O), ammonia (13NH3), and 82-rubidium (1). A new 18F-based radiotracer for PET myocardial perfusion imaging (Flurpiridaz F 18; Lantheus Medical Imaging) demonstrated excellent image quality in a phase I clinical trial (2) and better image quality, interpretative certainty, and diagnostic performance than SPECT in a phase 2 clinical trial (3). In particular, imaging with this 18F-based radiotracer generated higher-resolution cardiac images (4).
However, the high image resolution associated with an 18F-based tracer can be degraded by cardiac and respiratory motion during PET acquisition, leading to image blurring. If such motion is not corrected, the full-potential imaging resolution of Flurpiridaz F 18 may not be realized. We have previously developed a motion-frozen (MF) technique for myocardial perfusion imaging, which recovers resolution lost due to cardiac motion, improving contrast and quantitative diagnostic accuracy (5,6). This technique was also successfully used for automatic coregistration of SPECT with CT angiography (CTA) (7,8) and for contrast improvement in cardiac PET (9).
The aim of this study was to investigate the feasibility of myocardial perfusion imaging by Flurpiridaz F 18 PET with correction of both cardiac and respiratory motion. We hypothesized that the myocardial perfusion image quality for Flurpiridaz F 18 would be improved with dual-gated MF techniques beyond what is possible with the resolution recovery during reconstruction and MF techniques, which correct for cardiac motion only.
MATERIALS AND METHODS
Patients
The study population consisted of 16 patients (12 men and 4 women) recruited into a phase II clinical trial of Flurpiridaz F 18. The inclusion and exclusion criteria are given in the report of the clinical trial (3). Patients were instructed to discontinue caffeinated drinks and nitrate medications for 24 h and β-blockers and calcium antagonists for 48 h before testing. The study was approved by the institutional review board, and all patients provided written informed consent.
PET Acquisition
All patients underwent a rest–stress Flurpiridaz F 18 PET/CT scan as previously described (3). All images were acquired on a Siemens Biograph-64 TruePoint PET/CT scanner with the TrueV option. This 3-dimensional system consists of a 64-slice CT and a PET scanner with 4 rings of lutetium oxyorthosilicate detectors with a detector element dimension of 4 × 4 × 20 mm (10). The image plane spacing was 2 mm. The PET axial and transaxial fields of view were 216 and 605 mm, respectively. The coincidence time window and the energy window were 4.5 ns and 425–650 keV, respectively. The data were acquired in list-mode format. A full description of the system performance has been previously described (11).
After a 2.8-s standard topogram acquisition (120 kVp), patients underwent a CT scan for attenuation correction (CTAC) using the following parameters: spiral mode; slice thickness, 3.0 mm; total scan time, 3.36 s; pitch, 1.5; rotation time, 0.5 s; collimation, 24 × 1.2 mm; tube voltage, 120 kVp; and tube current, 11 mAs. The estimated radiation dose for this CTAC acquisition was 0.3 mSv. During the CTAC scan, the patients were instructed to perform normal end-expiration breath-hold. Rest perfusion images were acquired for 1 bed position around the heart (identical to the CT coverage of 210 mm) immediately on infusion of 85.1–159.1 MBq (2.3–4.3 mCi) of Flurpiridaz F 18 (depending on stressing protocol). Images were acquired for 15 min. Stress testing was performed either with a symptom-limited Bruce treadmill exercise protocol or with intravenous adenosine (140 μg/kg/min for 6 min). At near-maximal exercise, 277.5–388.5 MBq (7.5–10.5 mCi) of Flurpiridaz F 18 (depending on stressing protocol) were injected intravenously, after which treadmill exercise was continued at maximal workload for 1 min and at 1 stage lower for 2 additional minutes whenever possible. The 15-min perfusion PET acquisition was started at least 25 min after termination of exercise. For pharmaceutical stress with adenosine, Flurpiridaz F 18 was injected at the end of the third minute and a 15-min PET acquisition was started immediately. The interval between the rest and stress acquisitions was either 30 min when the patient was stressed with adenosine or 60 min when exercised stress was used. Twelve-lead electrocardiography was monitored continuously during stress testing.
The rest studies were acquired in cardiac-gating mode only whereas dual-gating (cardiac and respiratory) was used for all stress studies to minimize interference with the clinical trial. Therefore, we considered only stress studies here. The cardiac gating was performed with a 3-lead electrocardiography gating system inside the PET gantry. Respiratory gating was performed using a prototype pneumatic belt (Ivy Biomedical Inc.). This device sends the amplitude of the breathing motion to the PET system every millisecond. We used a modified acquisition control system to code the amplitude as tag-words inside the list-mode file at appropriate time positions. The amplitude-based respiratory gating has been shown to be more effective than the phase-based gating (12) with patients showing an irregular breathing pattern.
Reconstruction Protocols
List-mode files with embedded dual-gating signals were processed using an in-house-developed rebinning application. The extraction of the respiratory and cardiac signals allowed limitation of the reconstruction to specific sections of the cardiac or respiratory cycle. All images were reconstructed after a 5-min delay to allow for blood-pool clearance.
Rebinning
Two different reconstruction protocols were used for respiratory gating. The first one sorted the PET data list with an amplitude-based respiratory gating method, designed to keep the 35% of the PET counts in 1 optimal respiratory gate (ORG), in which the breathing motion was minimal (13).
In the second rebinning method, 4 separate list-mode files corresponding to 4 different breathing amplitudes were generated. We kept 20% of the total data in each of the 4 list-mode files corresponding to each respiratory gate, leaving out approximately 20% of events corresponding to the most extreme expiration and inspiration motion. The resulting 4 bins had equal time duration (but variable amplitude), which had the advantage of providing the same statistics in each bin. This initial rebinning of list-mode data allowed us to reconstruct dual-gated datasets.
Static and 8-bin electrocardiography-gated sinograms were then generated from all list-mode files. The sinograms from the original acquisition data list (no respiratory gating) included 5 min of the acquisition whereas the respiratory-gated and dual-gated sinograms were generated using 7 min of the acquisition to compensate for the reduced number of counts due to the respiratory rebinning, both starting 5 min after Flurpiridaz F 18 injection to allow for blood flow clearance.
Reconstruction
After the data list file sinograms were sorted, static summed images and 8-bin electrocardiography-gated images were reconstructed with HD•PET method (4 iterations and 14 subsets), which takes advantage of resolution recovery principles (14,15). For the fully dual-gated sinograms, 8-bin electrocardiography-gated by 4-bin respiratory-gated images were generated with HD•PET for a total of 32 bins. The reconstructed fully dual-gated datasets can be processed as 4 separate 8-bin electrocardiography-gated datasets, each of them at a different respiratory level. Figure 1 summarizes all the different reconstructions evaluated in this study. We chose to filter all patients with a 2-mm gaussian filter to maintain the same resolution and minimize noise suppression. The reconstruction matrix was 168 × 168 × 109 with a zoom of 2, and the pixel size was 2 × 2 × 2 mm. Scatter, decay, and random corrections were applied to the reconstructed images. Standard CT-based attenuation correction was used. When a misalignment was identified, a manual registration matrix with 3-dimensional translations was generated by an experienced technologist and applied before the final reconstruction process. Both the original PET/CT alignment and the alignment after manual registration were checked by an expert imaging physician. For the dual-gated datasets, 4 different PET/CT registration matrices were generated: 1 for each respiratory phase. Manual registration between PET and CT was applied before reconstruction when a misalignment was identified.
Overview of reconstructed datasets. MF = motion frozen; ORG = optimal respiratory gating; dual MF = cardiac and respiratory MF processing. After postprocessing, all final series are static.
Postprocessing Applied to Reconstructed Image
MF Technique
MF processing tracks the motion of the left ventricle (LV) in cardiac-gated images and then adjusts all cardiac phases to end-diastolic position, resulting in an image free of cardiac motion. Motion freezing is accomplished by motion tracking of the detected LV endo- and epicardial borders (5,9). The algorithm generates 3-dimensional displacement vectors between each cardiac phase and uses those vectors to shift all the counts to the end-diastolic phase. The result is an image free of cardiac motion blurring (i.e., with the same spatial resolution as the single cardiac phase image) but with greater counts and similar noise as compared with static summed images. This algorithm has been previously described in detail and validated for SPECT versus invasive angiography (5).
Dual-MF Technique
The dual-MF technique was applied in 2 consecutive steps. Starting from the 4 separate 8-bin cardiac-gated sets (each corresponding to a different respiratory phase), the MF technique previously described was first applied to shift 8 cardiac bins to end-diastolic position, separately for each respiratory bin. This resulted in 4 cardiac motion–corrected respiratory bins, each adjusted to a separate end-diastolic position. Subsequently, to compensate for the respiratory motion, we applied rigid transformation (6 parameters including 3 translations and 3 rotations), accounting for the difference between reference and target frames by performing consecutive registrations. For the reference frame, we selected the first frame, representing the position during end-expiration. We applied a bilateral filter (16) to remove noise while preserving edge information and generated a binary mask to localize the region of interest (myocardium). Consequently, automated frame-to-frame rigid registration of frames 2–4 to frame 1 (end-expiration) was performed using a similarity measure of normalized cross correlation and gradient descent minimization, in which sample voxels are selected from the mask region for computational efficiency. This algorithm was previously developed by our group for gross patient motion correction of PET data (17). The overview of the dual-respiratory/cardiac-MF processing is shown in Figure 2.
Illustration of dual-respiratory/cardiac-MF technique: 1. Cardiac-MF processing separately in each respiratory phase (contoured red bins) to end-diastole phase (solid red bins). 2. Respiratory MF processing of cardiac end-diastole bins (contoured blue bins) to end-inspiration bin (solid blue bin). 3. Final motion-free image in end-diastole/end-inspiration (solid green bin). ECG = electrocardiography.
Image Processing and Analysis
Short-axis reorientation and automatic contouring of the LV in static and gated images were done automatically using QPET software (Cedars Sinai cardiac package) for which the inputs were the transverse slices of the reconstructed PET image (18). MF and dual-MF image processing techniques were then applied to the gated datasets (Fig. 1). The myocardial wall segment analysis was performed using the 17-segment American Heart Association model (19).
Myocardial Wall Thickness
Using the LV coordinates automatically detected by QPET, we estimated the wall thickness defined as the full width half maximum of a profile taken on the images at 3 different levels of the long axis (basal, mid, and apical) and with 2 orientations (septal-lateral and inferior-superior) for a total of 6 profiles per dataset (Fig. 3).
Illustration of wall thickness calculation. Wall thickness is defined as the full width half maximum of a profile taken on images at 3 different levels of long axis (basal, mid, and apical) and with 2 orientations (septal–lateral and inferior–superior) for a total of 6 profiles per dataset. Max = maximum
Myocardium-to-Blood Contrast and Contrast-to-Noise Ratio (CNR)
Volumes of interest over the LV and the blood pool (inside the LV cavity) were automatically derived on the basis of the QPET segmentation of the LV. Their dimensions were significantly larger than the spatial resolution of the images. We then calculated the contrast between the blood pool and the myocardial wall, and the CNR as follows:where SD is the SD of the blood pool. Contrast and CNR were calculated globally and for each of the 17 segments as defined by American Heart Association, from static summed images and MF processed images.
Statistical Analysis
All continuous variables are expressed as mean ± SD. Paired t tests were used to compare differences in paired continuous data, and the McNemar test was used to compare differences in paired discrete data. For unpaired continuous data, 1-way ANOVA was used. All statistical tests were 2-tailed, and a P value of less than 0.05 was considered significant.
RESULTS
Patient demographic characteristics are shown in Table 1.
Patient Demographic and Clinical Characteristics (n = 16)
Myocardial Wall Thickness
The measured wall thickness was significantly different between all datasets. We found that the wall thickness was 13.9 ± 2.8 mm for ungated, 12.1 ± 2.7 mm for cardiac-MF, 13.1 ± 2.9 mm for ORG, and 11.6 ± 1.9 mm for dual-MF images (Table 2). The dual-MF technique resulted in a decreased wall thickness for 94%, 77%, and 91% of the measurements as compared with ungated, cardiac-MF, and ORG data, respectively. For comparison, we also measured the wall thickness for the end-diastole end-expiration phase of the dual-gated datasets, which was 10.7 ± 2.3 mm.
Myocardial Wall Thickness (in mm) Measured on Profiles Taken at Basal, Mid, and Apical Levels in Septal–Lateral and in Superior–Inferior Directions
Myocardium-to-Blood Contrast and CNR
The myocardium-to-blood contrast significantly changed between the different gating and processing techniques as shown in Figure 4. Figure 5 provides CNR results on a per-patient/per-method basis. Examples of improvements in terms of visual image quality with the different gating and processing techniques can be seen in Figures 6 and 7.
Myocardial-to-blood contrast and CNR for ungated, cardiac MF with no respiratory gating, ORG, and dual (respiratory and cardiac)-MF images. P < 0.05 for all except ungated vs. ORG.
CNR for ungated, cardiac MF with no respiratory gating, ORG, and dual (respiratory and cardiac)-MF images, for each of the 16 patients.
A 42-y-old female patient with low likelihood of coronary artery disease (weight, 86.2 kg [190 lbs]; body mass index, 34.7) injected with 284.9 MBq (7.7 mCi) of Flurpiridaz F 18 during exercise stress and acquired in dual (cardiac and respiratory) mode. Short-, vertical-, and horizontal-axis views of adenosine stress images. 1 phase = 1 cardiac gate in end-inspiration; dual MF = dual (cardiac/respiratory) MF; MF = cardiac MF no respiratory gating.
A 67-y-old male patient with prior myocardial infarction (weight, 88.5 kg [195 lbs]; body mass index, 25.7) injected with 355.2 MBq (9.6 mCi) of Flurpiridaz F 18 during adenosine stress and acquired in dual (cardiac and respiratory) mode. Short-, vertical-, and horizontal-axis views of adenosine stress images. 1 phase = 1 cardiac gate in end-inspiration; dual MF = dual (cardiac/respiratory) MF; MF = cardiac MF no respiratory gating.
DISCUSSION
We have shown that with postprocessing techniques, dual-gating is possible with Flurpiridaz F 18 myocardial perfusion imaging and that this approach provides almost motion-free images with low noise and greater apparent resolution. These techniques therefore hold major promise to improve the sensitivity and resolution of myocardial perfusion imaging studies and cardiac PET imaging in general.
We corrected for both cardiac and respiratory motion that would usually blur the typical static summed perfusion images. We first investigated the feasibility of decreasing the effect of the respiratory motion with an optimal respiratory gating technique that keeps only a narrow range of breathing amplitude (35%) around the end-expiration. van Elmpt et al. have previously applied this approach in PET-based radiation therapy planning, though they used different methods to measure and record respiration (13). Another recent publication also considered the same approach (20). This technique eliminates about 65% of the counts and uses 35% of the remaining counts in the quiescent phase of the breathing to provide images almost free of respiratory motion.
Although in the original Lantheus protocol, 15 min of PET emission data (10 min acquisition starting 5 min after injection to allow clearance from blood pool) were collected to allow for the range of scanners used in the trial, our center had established that a 5 min perfusion acquisition provided clinically robust image quality on the 4-ring Siemens Biograph PET/CT scanner. We increased the acquisition time from 5 to 7 min for respiratory- and dual-gated datasets to allow similar contrast-to-noise levels for all data. Further, a shorter imaging time has the added benefit of reducing the potential for patient motion, which could confound the effects of respiratory and cardiac motion.
Because the ORG method wastes about 65% of the counts, we looked into the feasibility of full dual-gated imaging where the MF technique is applied for both respiratory and cardiac motion correction. The CNR in the dual-MF images using 7 min of acquisition was higher than in the ORG images and consequently the increase in acquisition time may not be needed for the dual-MF technique to keep the noise level similar to the static summed perfusion images.
Correcting for both cardiac and respiratory motion leads to a decreased myocardial wall thickness. The wall thickness has decreased by about 17% for the dual-MF technique as compared with the standard ungated method currently used for perfusion imaging. These results are closer to average thickness as reported by MR imaging techniques (10.3 ± 1.2 mm at end-diastole) (21). The myocardium-to-blood contrast increased more than 59% and CNR increased more than 89% for the dual-MF technique as compared with the reference ungated technique. Even when compared with a cardiac-only MF method, the dual-MF method allowed a 37% increase in contrast and 19% increase in CNR. There were 4 cases, however (Fig. 5), in which the dual-gated technique did not improve CNR over the cardiac-MF method, despite improving contrast. This noise increase is likely due to difficulties in convergence of the iterative reconstruction for low-count individual double-gated bins. More robust reconstruction techniques (22) could be used in future studies to reduce this effect. The ungated images in this study were reconstructed with the state-of-the-art technique incorporating resolution recovery principles, which has been shown to have better quality than regular iterative reconstruction for cardiac PET (15).
This is the first attempt to optimize the reconstruction of myocardial perfusion Flurpiridaz F 18 images with respect to cardiac and respiratory motion. Dual-gating (cardiac/respiratory) (23–26) and preliminary methods for dual motion correction for phantom data and 18F-FDG have been reported (27–29) but have not yet been applied to myocardial perfusion imaging and did not use the advanced reconstruction with resolution recovery. Clinical implementation of the dual-gating technique will likely lead to the improved clinical utility of the Flurpiridaz F 18 images but nevertheless adds additional technical complexity. Although the amplitude-based respiratory gating systems are now routinely used for oncologic applications, the registration methods and the 2-step list-mode rebinning used in this study are not yet widely used and may require new streamlined software before they can be routinely applied in clinical practice. Further, in this study we ignored the effect of misalignment of the attenuation scans with individual respiratory and cardiac gates of the PET emission data. Ideally, attenuation maps should be corrected for both cardiac and respiratory motion before the reconstruction of individual bins. However, this will require the use of more complex methods not yet available clinically (30).
Significant improvements in spatial resolution and contrast resolution are possible with an 18F-labeled tracer such as Flurpiridaz F 18, because its positron range is short (0.54 mm). Therefore, the correction of cardiac and respiratory motion may be of particular importance in this case, allowing the full benefit of improved perfusion image quality for this tracer. Motion as small as 5 mm can significantly affect the quantitative perfusion analysis, and the analysis of subtle changes between stress and rest myocardial perfusion PET may be affected by changes as small as 5%–10% in the reduction of the apparent maximal activity (31). In our previous work applying MF techniques to SPECT, correcting for cardiac motion only, we were able to improve the diagnostic accuracy of the techniques (5,6).
There was no gold standard for wall thickness but end-diastole end-expiration dual-gated phase thickness was measured and compared with dual-MF results. Although apparent contrast and CNR have increased, the clinical significance of these improvements and in particular the potential improvement of perfusion abnormality detection will need to be determined in future work. Our cohort of patients had a limited range of body mass index, typical of our patient population. Our sample size was small (n = 16), and further studies with a larger number of patients are warranted to assess more fully the diagnostic value of this technique.
CONCLUSION
In this work, we showed that dual-gating with cardiac and respiratory motion correction was feasible with Flurpiridaz F 18. Dual-gated MF imaging provides high-quality motion-free perfusion images with improved contrast and image resolution.
DISCLOSURE
The costs of publication of this article were defrayed in part by the payment of page charges. Therefore, and solely to indicate this fact, this article is hereby marked “advertisement” in accordance with 18 USC section 1734. Ludovic Le Meunier is an employee of Siemens Medical Systems. Joel L. Lazewatsky is an employee of Lantheus Medical Imaging. Piotr J. Slomka and Daniel S. Berman have received research funding from Siemens. Daniel S. Berman has received research support from Lantheus. No other potential conflict of interest relevant to this article was reported.
Acknowledgments
We thank Jim Hamill for his help with the optimal respiratory gating method.