Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
LetterLetters to the Editor

Injected Dose in Pediatric PET

Charles C. Watson
Journal of Nuclear Medicine October 2010, 51 (10) 1657; DOI: https://doi.org/10.2967/jnumed.110.078493
Charles C. Watson
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Info & Metrics
  • PDF
Loading

TO THE EDITOR: In their recent paper, Accorsi et al. (1) provide a detailed quantitative analysis of the efficacy of pediatric patient injected dose in 18F-FDG PET. For their analysis, they have defined a quantitative metric of data quality, the “noise-equivalent count density” (NECD), as a surrogate for clinical image quality. NECD is the total noise-equivalent counts (NEC) in the measured data divided by an estimate of the volume of the patient within the field of view of the scanner. I believe that in general the use of such a data quality measure as a proxy for imaging performance is a valuable approach for guiding scanning protocols. As the authors mention, NEC has been widely used for this purpose in other contexts. I would like to point out, however, that NECD is a metric qualitatively different from NEC and that its use leads to dose recommendations significantly different from what would have been derived from using NEC as the metric.

The motivation for characterizing data quality in terms of NECD rather than NEC is not spelled out in the paper. My tentative assumption is that the goal is to approximately equalize the signal-to-noise ratio (SNR) per unit image volume. Although there are many confounding issues, intuitively one might suppose that if the number of (noise-equivalent) counts per unit volume were the same for 2 patients, then the SNR (mean divided by the SD of the noise) for volumes of interest of the same size in the 2 images would likely be similar on average. Thus, pediatric and adult scans performed with similar NECDs might be expected to have similar SNRs in anatomically matched regions of equal volume—which may seem to be a valid criterion for equivalent image quality.

I would question such an assumption, however. It seems to me that tissue volumes of clinical interest are in general more likely to scale with the size (volume) of the patient. Many disease processes imaged in PET tend to be primarily associated with a specific organ (such as the heart), and children have smaller organs than adults. And, although I have no hard evidence supporting this hypothesis, it also seems plausible that the physiologic impact of a cancerous neoplasm would scale at least to some extent with the size of the body and, thus, that the detection of smaller tumors in smaller bodies would be required for equivalent efficacy.

Roughly speaking, an equalized NEC-based regimen would tend to equalize SNR per organ rather than per volume. Therefore, to the extent that clinically meaningful volumes of interest also scale with organ or body size, I believe that NEC may be a better predictor of image quality than is NECD. I therefore find the authors' statement (page 296) that “…the current [weight-based] injection regimen produces better image quality in lighter patients…” doubtful. It seems more probable to me that the recommended constant-NECD regimen would produce poorer image quality in lighter patients.

Certainly, minimizing unnecessary injected dose in pediatric PET is an extremely important goal, and the authors are to be commended for their careful analytic approach to it. The equalized NECD regimen they propose would in fact result in much smaller doses (approximately proportional to the patient's weight) than would a similar strategy based on NEC. My concern is that this regimen carries the risk that injected dose could be reduced to the point that the clinical value of the PET scan does not, in the end, justify the deleterious effects of the radiation absorbed by the patient. I believe it would be most prudent to maintain NEC as high as reasonably possible until there is solid evidence that diagnostic image quality is adequately estimated by NECD.

A second fundamental assumption underlying this paper is that the NECD observed for a 3-min scan of a 70-kg adult on the Gemini 16 PET/CT scanner (Philips Health Care) with a 363-MBq injected dose is clinically good enough and that any excess NECD can be traded off for lower dose or shorter scanning time. Yet I have never seen a convincing clinical study quantifying how many noise-equivalent counts are enough in PET. I believe that any PET image can be improved by additional valid data and, thus, that it is in the patient's best interest, when scan time is not limited by radiopharmaceutical redistribution or decay, to scan every patient for as long as is practical, regardless of dose. Further, for 18F-FDG imaging I see no reason to limit the injected dose to less than that estimated to achieve about 90% of peak noise equivalent count rate unless the health risk of the absorbed radiation outweighs the likely clinical benefit of the scan. Unfortunately, as we know, this criterion is exceedingly difficult to quantify. Yet I believe that the definition of a more efficacious dose regimen can come only from a better understanding of this trade-off.

  • © 2010 by Society of Nuclear Medicine

REFERENCE

  1. 1.↵
    1. Accorsi R,
    2. Karp JS,
    3. Surti S
    . Improved dose regimen in pediatric PET. J Nucl Med. 2010;51:293–300.
    OpenUrlAbstract/FREE Full Text
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 51 (10)
Journal of Nuclear Medicine
Vol. 51, Issue 10
October 1, 2010
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Injected Dose in Pediatric PET
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Injected Dose in Pediatric PET
Charles C. Watson
Journal of Nuclear Medicine Oct 2010, 51 (10) 1657; DOI: 10.2967/jnumed.110.078493

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Injected Dose in Pediatric PET
Charles C. Watson
Journal of Nuclear Medicine Oct 2010, 51 (10) 1657; DOI: 10.2967/jnumed.110.078493
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • REFERENCE
  • Info & Metrics
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • Evaluation of Optimal Acquisition Duration or Injected Activity for Pediatric 18F-FDG PET/CT
  • Google Scholar

More in this TOC Section

  • Business Model Beats Science and Logic: Dosimetry and Paucity of Its Use
  • Determining PSMA-617 Mass and Molar Activity in Pluvicto Doses
  • The Value of Functional PET in Quantifying Neurotransmitter Dynamics
Show more Letters to the Editor

Similar Articles

SNMMI

© 2025 SNMMI

Powered by HighWire