JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Right arrow Help viewing high resolution images
Right arrow Return to article
Click on image to view larger version.


Figure 9


FIGURE 9.  Two-dimensional PET sinograms and reconstructed transverse images through hot-sphere 18F-filled phantom: sinogram without any perceptible artifacts (A) and corresponding transverse image (B); sinogram with blank diagonal band, suggestive of faulty detector block (C) and corresponding transverse image (D). Despite obvious defect in sinogram C, reconstructed images B and D are virtually indistinguishable. (Courtesy of Dr. Osama Mawlawi, M.D. Anderson Cancer Center, Houston, TX.)