FIGURE 6. (A) Severe ring, or bull's-eye, artifact in transverse SPECT image through 20-cm diameter 99mTc-filled cylinder phantom resulting from gross
-camera nonuniformity (i.e., IU
10%). (B) Appearance of such artifact (arrow) in clinical SPECT image, transverse image from 99mTc-sulfur colloid liver–spleen study. (C) In contrast to A, this is acceptable transverse SPECT image through 99mTc-filled cylinder phantom. No perceptible ring, or bull's-eye, artifact is demonstrated; IU at time of this acquisition,
2%, was well within tolerance. In cylinder phantom images, A and B, analytic postprocessing attenuation correction was applied.