Abstract
Localization of regions with increased uptake of radiotracer in small-animal SPECT is greatly facilitated when using coregistration with anatomic images of the same animal. As MRI has several advantages compared with CT (soft-tissue contrast and lack of ionizing radiation) we developed a SPECT/low-field MRI hybrid device for small-animal imaging. Methods: A small-animal single-pinhole γ-camera (pinhole, 1.5 mm in diameter and 12 cm in focal length) adjacent to a dedicated low-field (0.1 T) small MR imager (imaging volume, 10 × 10 × 6 cm3) was used. The animal was placed in a warmed nonmagnetic polymethyl methacrylate imaging cell for MR acquisition, which was followed immediately by SPECT after translation of the imaging cell from one modality to the other. 3-Dimensional T1-weighted sequences were used for MRI. Phantom studies enabled verification of a low attenuation (10%) for 99mTc and 201Tl and a very slight increase in Compton scattering due to the radiofrequency coil and polymethyl methacrylate imaging cell. Results: SPECT/MRI data acquisition and image coregistration of selected examples using different radiotracers for lungs, kidneys, and brain were obtained in 3 nude mice with isotropic spatial resolutions of 0.5 × 0.5 × 0.5 mm3 for MRI and 1 × 1 × 1 mm3 for SPECT. The total acquisition time for combined SPECT and MRI lasted 1 h 45 min. Conclusion: A low-magnetic-field strength of 0.1 T is a simple and useful solution for a small-animal dual-imaging device combining pinhole SPECT with the adjacent MR imager.
Among small-animal imaging techniques, SPECT provides a unique possibility to follow and measure in vivo, and noninvasively, the biodistribution of a 10−9 molar concentration of a wide range of radiolabeled biomolecules commonly available in nuclear medicine departments (1). However, one essential drawback in SPECT images is the lack of anatomic references related to the tissue uptake of tracer. Therefore, localization of regions with increased uptake of radiotracer is greatly facilitated when using coregistration of SPECT with anatomic images of the same animal, from CT or MRI. Over the last few years, SPECT/CT dual modality has been widely proposed by manufacturers but MRI presents specific advantages compared with CT, including lack of ionizing radiation, high soft-tissue contrast, and sensitivity to tissue alterations evidenced by specific imaging sequences (2,3). Yet, compared with CT, the use of MRI for coregistration of both functional and structural information has, to our knowledge, essentially served to demonstrate the potential interest of using coregistration of images after data acquisition in separate nuclear and MR rooms. However, as experienced for rat and mice, the strategy of pinhole SPECT followed by MRI in a clinical scanner is a long and complicated task (4–7), requiring a careful transfer of the animal in a specially designed bed equipped with multimodality fiducial markers helping in the coregistration of images (4,5). In addition, separate dual-modality scans (followed by image fusion) are also potential sources of misalignment of images due to uncontrolled movements and displacements of tissues and organs. Moreover, multiple anesthesia sessions are needed when SPECT and MRI devices are located in separate places, involving the risk of different biologic responses to anesthesia (3). After early rodent MRI works done on clinical low-field systems operating at 0.35 T (8), small-animal MRI is now performed exclusively at very high magnetic fields operating at values up to 14 T, in a desire to achieve high anatomic resolution benefiting from a higher signal-to-noise ratio but leading to field susceptibility artifacts and preventing close proximity of the SPECT scanner (2,9). As for PET/MRI, technical limitations for dual SPECT/MRI systems are therefore primarily due to complex physical interactions between the high magnetic field of the MR device and nuclear instrumentation. Two strategies are then possible for the design of a SPECT/MRI dual-modality system for small-animal imaging. The first strategy is the development of high-magnetic-field MR-compatible SPECT inserts as was recently proposed at 7 T (10). However, a similar approach for MR-compatible PET inserts resulted in the costly design of small-animal prototypes for simultaneous dual-modality devices working for MRI at 4.7 and 7 T (11,12). The second strategy is the use of a low magnetic field for MRI that could have the combined advantages of avoiding field strength imaging artifacts and interactions with the SPECT instrumentation while providing sufficient pixel resolution for an appropriate routine solution for a small-animal dual-imaging device that would minimize time and distance between the scans. After a previous proof-of-concept study (13), in this article we propose to describe, illustrate, and discuss our practical experience in small-animal sequential SPECT/MR image acquisition and fusion using a coaxially back-to-back configuration of a pinhole SPECT with a 0.1-T MRI device.
MATERIALS AND METHODS
Animal Holder and Anesthesia
All animal experiments were conducted in compliance with the French guidelines for the care and use of research animals (authorization 67-482-20). Mice were maintained under gaseous anesthesia (1%–1.5% isoflurane in air) in an air-warmed (40°C), nonmagnetic technical cell dedicated to small-animal imaging and adapted for both SPECT and MRI devices (CTI; Minerve) (Fig. 1). The imaging cell isolates the animal during image acquisition of both modalities. The use of the imaging cell equipped with a single radiofrequency (RF) transmit-and-receive coil allows an easy and immediate transfer of the cell from one imaging system to the other while keeping the animal in the same position. Therefore, fiducial markers in the imaging cell were avoided in our experiments.
(A) General view of hybrid imaging system with pinhole SPECT (a) adjacent on a same-axis configuration to low-field MR imager (b) and close to the anesthesia equipment (c). Imaging cell and animal (3) are translated along the imaging axis (2) from the MRI device to a single-detector pinhole SPECT device (1). Illustration shows orientation of the vertical B0 main magnetic field of the MR magnet and corresponding power supply (4). (B) Close-up view of the nonmagnetic imaging cell: solenoidal whole-body RF transmit-and-receive coil (1); pickup coil (2) outside the isolating heated polymethyl methacrylate chamber (3); anesthetic gases are delivered to animal through a dedicated mask (4).
SPECT System and Image Reconstruction
For SPECT we used a small-animal dedicated rotating single-head γ-camera with a 1.5-mm-diameter tungsten pinhole collimator and 12 cm in focal distance (Gaede). For small-animal SPECT image acquisition, a 20% window centered on the 140-keV photopeak was used, and 60 projections of 1 min over an arch of 180° in a 128 × 128 format were obtained using a 4-cm radius of rotation around the imaging cell. Images were reconstructed with an in-house efficient algebraic cone beam reconstruction method based on an algebraic reconstruction technique (ART) algorithm that takes into account the variation of the radius of rotation and is able to achieve an isotropic 1 × 1 × 1 mm3 spatial resolution (efficient algebraic reconstruction [EAR] method (14)). The size of the isotropic reconstructed voxel was 0.47 × 0.47 × 0.47 mm3. The cost of the complete pinhole SPECT apparatus was around $50,000.
MRI System and Design of RF Coils
MRI was performed with a small water-cooled low-field (main magnetic field strength B0 = 0.1 T oriented vertically) open electromagnet (Bouhnik SAS) that was previously described in detail (15). It has a homogeneous (relative variation of magnetic field strength ΔB/B0 = 5.10−6) imaging volume of 10 × 10 × 6 cm3 and maximum encoding gradients' strength of 20 mT/m that can be reached in <0.5 ms. RF transmit-and-receive solenoidal coils (for whole-body or brain images) were made by adjacent loops of a 0.7-mm-diameter copper wire rounded on a thin (0.2 mm) plastic cylinder to minimize attenuation and scattering of γ-rays during SPECT. The animal was placed from the beginning in the imaging cell as illustrated in Figure 1 and kept in the same position during both acquisitions. RF pulses are transmitted to the solenoidal coil by a pickup copper coil rounded outside the imaging cell (as illustrated in Fig. 1), which also received the encoded MR signal. A T1-weighted 3-dimensional (3D) sequence was developed on a SMIS (Guilford, U.K.) console to achieve isotropic spatial resolution of 0.5 × 0.5 × 0.5 mm3. T1-weighted isotropic acquisitions were obtained in 42 min with echo time (TE) = 7 ms, repetition time (TR) = 23 ms, flip angle (FA) = 60°, 128 × 64 × 48 or 256 × 64 × 48 matrix, number of excitations (NEX) = 36, and field of view (FOV) = 64 × 32 × 32 mm3 or 128 × 32 × 32 mm3. The cost of the complete MR equipment was around $200,000.
Combined Sequential SPECT/MRI Dual Modality
As the sizes of our 2 imaging devices (SPECT: 103-cm length by 80-cm width and 140-cm height; low-field MRI: 72-cm length by 72-cm width and 118-cm height) fit in a single room, and as the 5-Gauss line of the magnetic field is located at 15 cm from the sides of the magnet, the pinhole SPECT camera was placed directly close and coaxially to the MR magnet as well as the anesthesia equipment without any adverse effects between them (as illustrated in Fig. 1). After MR acquisition, the pickup coil was removed, and the imaging cell containing the animal and the RF coil was manually translated, on the common z-axis, from the center of the magnet to the pinhole SPECT system and docked in <1 min.
Phantom Studies
The influence of the RF coil and the polymethyl methacrylate tube wall of the imaging cell on attenuation and scattering of γ-rays was tested for 2 different isotopes in SPECT projection images of a simple tube filled with 37 MBq of a solution of 99mTc and 201Tl, respectively. Windows of 20% centered on 140 keV for the 99mTc and on 71 and 167 keV for the 201Tl, respectively, were used. A second phantom consisting in 3 tubes of different diameters, described in Figure 2A and placed inside the imaging cell, was used to set out the centering errors consecutive to the manual translation of the imaging chamber from one imaging modality to the other. Only one tube of this phantom was filled with 37 MBq of 99mTcO42− for SPECT, whereas the others were filled with water and water doped with CuSO4 (2 g/L) for MR T1 contrast enhancement. The T1-weighted imaging sequence defined earlier was used for MRI. For the SPECT study, a 4-cm radius of rotation and a 1.5-mm pinhole diameter were used, and 32 projections of 1 min were acquired over a circular 180° arch with a 128 × 128 image matrix. Pinhole SPECT images were reconstructed using the EAR method. Then, using AMIDE software (16), rigid coregistration parameters for both datasets were obtained, after identification of common features in each image, by orthogonal x-, y-, and common z-axis translations only and used to correct the centering errors for the in vivo experiments.
(A) Diagram and coregistrated MR and SPECT images of phantom made of 3 tubes (a, b, and c) filled with water. One inner tube is filled with 99mTcO42− (b), whereas the other one is doped with CuSO4 (c). (B) Comparison of 99mTc and 201Tl spectra without (yellow) and with (green) RF coil and imaging cell. A y-logarithmic scale was used to enhance the slight differences observed in Compton diffusion for both isotopes. (C) Activity line profile of the single-tube phantom filled with 99mTcO42− (left) and the same tube rounded by the RF coil inside the imaging cell (right). Low γ-rays' attenuation of 10% due to RF coil (a) and imaging cell wall is clearly visible.
Animal Population, MR Contrast Agent, SPECT Radiopharmaceuticals, and Coregistration of Images
Three adult nude mice, each weighting 25 ± 1 g (mean ± SD), were studied using different radiotracers to label different organs and tissues. Pinhole SPECT immediately followed MR acquisition in all experiments. Radiopharmaceuticals with a volume of 0.2 mL were injected in the tail vein 10 min before beginning the MR procedure. Corresponding activities were respectively 108 MBq of 99mTc-albumin (Vasculocis; Cis-Bio International) for lung perfusion, 98 MBq of 99mTc-dimercaptosuccinic acid (Renocis; Cis-Bio International) for kidney, and 120 MBq 99mTc-hexamethylpropyleneamine oxime (Ceretec; GE Healthcare) for brain perfusion. Intraperitoneal administration of 1 mL of 0.1 mmol of gadoteric acid (Dotarem; Guerbet) was used before intravenous injection of radiopharmaceutical for MR tissue contrast enhancement. In all examples, rigid coregistration of images was obtained applying the set of correction values from the 3-tube phantom using the AMIDE software (16). The total acquisition time for SPECT and T1-weighted MR images in mouse lasted 1 h 45 min.
RESULTS
Phantom Studies
Comparison of total counts in SPECT projection images without and with RF coil and imaging cell show a relative low attenuation of 9.7% for 99mTc and 10.5% for 201Tl. The 2-mm wall thickness of the imaging tube cell contributed to 5% of the total attenuation. In addition, a very slight increase in Compton scattering was observed in both spectra due to the RF coil and wall thickness of the imaging cell tube but without any influence on the counting windows (Fig. 2). Figure 2 also shows the coregistration of SPECT and MR images of the phantom consisting of 3 tubes.
Animal Studies
Figure 3 shows selected slices of MR T1-weighted whole-body mouse images in which anatomic structures, such as heart cavities, gallbladder, or vertebras, can be easily described. Corrections of off-center misalignment of the imaging cell using the 3-tube phantom data were applied before coregistration of SPECT/MRI mouse images. Anatomic references were then chosen on SPECT and MRI datasets for each experiment, and a mean matching coregistration error of 0.93 mm was measured using AMIDE software. Figures 4, 5, and 6 illustrate the coregistration of MR and SPECT images in the 3 orthogonal planes for the different radiopharmaceuticals used in this study. The matching of lung, kidney, and brain functional SPECT images on MRI anatomic images demonstrates the practical interest of this dual-modality system.
Example of selected sagittal and coronal whole-body low-field MRI slices of a nude mouse from the whole set of isotropic (0.5 × 0.5 × 0.5 mm3) anatomic T1-weighted images.
(A) T1-weighted low-field MRI coronal slice (in gray) showing lack of signal in lungs and landmark placed manually at apex of right lung. (B) Corresponding SPECT image of lung perfusion (in color) and landmark placed manually at apex of right lung. (C) Fusion of both images where anatomic MRI-based and corresponding functional SPECT landmarks are shown on this particular slice assessment of coregistration error calculated as mean distance between 4 coregistered landmarks per complete image dataset.
Selected coregistered orthogonal slices of pinhole SPECT (99mTc-dimercaptosuccinic acid [99mTc-DMSA] kidney uptake in color) and low-field MRI (soft-tissue anatomic information in gray) in nude mouse.
Selected coregistered orthogonal slices of pinhole SPECT (brain perfusion in color) and low-field MRI (soft-tissue anatomic information in gray) in nude mouse. HMPAO = hexamethylpropyleneamine oxime.
DISCUSSION
Spatial Resolution of SPECT and MRI
Pinhole SPECT is an intrinsic 3D imaging method with isotropic submillimeter-to-millimeter spatial resolution currently achieved in small animals (1,17–22). In our experiments, an isotropic spatial resolution of 1 mm was obtained by pinhole SPECT with a reconstructed voxel of 0.47 × 0.47 × 0.47 mm3 (14,23). It is also obvious that isotropic voxel MR images are recommended for optimal coregistration with SPECT. We were able to achieve 0.5 × 0.5 × 0.5 mm3 voxels for MRI. They can be compared with 0.3 × 0.3 × 0.3 mm3 voxels obtained in mice in separate studies and machines in the only study dealing with MR isotropic voxels, even though this study did not mention the magnetic field strength used (6). Current MRI strategies for small animals are more frequently based on 2-dimensional (2D) multislice acquisition instead of time-consuming isotropic volume imaging sequences. In-plane pixel resolution ranging from 0.15 to 0.25 mm and slice thickness ranging from 0.7 to 2 mm were obtained using clinical scanners at 1.5 T for separate SPECT and MRI rat brain studies (4,5,7). Also, for comparison, a 0.8 × 0.8 × 0.8 mm3 voxel has been used for a whole-body mouse MR acquisition in separate PET and MRI studies using a 4.7-T magnetic field (24). However, SPECT/MRI fusion studies based on nonisotropic voxels for both modalities lack precision for image coregistration and fusion because the MRI slice thickness differs from the pixel size, even if the MR in-plane pixel resolution is improved. Our results show that the voxel size achieved at 0.1 T is adapted for an easy routine coregistration for SPECT functional images with anatomic MR references. As physiologic movements of the animal and tissues due to circulation, respiration, and fluid flow cannot be systematically controlled, it also seems unrealistic to aim for MR spatial in-plane resolution below 0.2 × 0.2 mm2 (25,26). Compared with high and very high magnetic fields, dealing with the relative voxel sizes achieved by each of the combined technologies also does not make sense to merge SPECT data, whose spatial resolutions are on the order of 1 mm3, with MR spatial resolutions of 0.001 mm3. Of course, the major, but almost only, disadvantage of a low magnetic field compared with a high field (proton spectroscopic information notwithstanding) is the reduction in the signal-to-noise (S/N) ratio. However, improvements with regard to the S/N ratio and resolution at low magnetic field are possible by increasing the gradient strength, which increases both voxel resolution and signal because it shortens the echo time. The use of superconducting or copper-cooled RF coils also allows an increase in the S/N ratio but remains questionable because of the necessity of heating the imaging cell at the same time (27,28).
Total Acquisition Time for SPECT/MRI
In our experience, the total acquisition time of the complete set of SPECT/MR images was <2 h. As we used the technical imaging cell for both modalities, the manual transfer of the animal from SPECT to MRI was effective in about 1 min, but this translation could also be done automatically as SPECT and MRI components are adjacent to one another on the same imaging axis. The acquisition time of SPECT images alone could also be optimized in view of recent technical improvements—such as multiple heads' NaI crystal detectors, multiple pinholes, or multiple semiconductor detectors—which can increase the sensitivity of γ-photon detection (18–22). In our experience, total acquisition times with low-field MRI were identical to the 43 min in the experience of isotropic voxel high-field MRI used for PET/MRI coregistration images in mice (24). Obviously, multislice 2D MR acquisitions are faster (about 15 min, for instance, at 1.5 T (4,5)) but, as mentioned earlier, are not well adapted for adequate coregistration of images because the MRI voxel needs to be resized to match the isotropic SPECT voxel size. Moreover, 3D MRI sequences had a better S/N ratio relative to the acquisition time than corresponding 2D multislice sequences, which is an advantage when using a low magnetic field strength but at the expense of longer acquisition times (29).
Imaging Cell for Both Online Sequential Modalities
Keeping the animal motionless was achieved by the use of the imaging cell, allowing easy, safe, and rapid transport of the anesthetized animal from one imager to another—therefore, minimizing the delay between both sets of image acquisitions. As the animal remains anesthetized in the imaging cell from the beginning of the sequential imaging modalities, the lack of fiducial markers inside the imaging cell was not critical for coregistration of SPECT/MR images in our experiments. It is true that we found a mean coregistration error corresponding to the value of the spatial resolution of the pinhole SPECT, which is the less-resolved modality. However, use of a nonrigid spatial transformation matrix based on grid phantoms and fiducial markers could be easily mounted on the imaging cell to facilitate improvement in the accuracy of images' coregistration as done by Chow et al. (30). Finally, the use of the imaging cell enables registration of both sets of images during the same anesthesia session, thus avoiding various biologic responses due to different doses and types of anesthesia as could be the case in completely separate acquisitions (3).
Small-Animal Control and Survey During SPECT/MRI Dual Imaging
A specific advantage of the combined pinhole SPECT and dedicated low-field MR imager used in this study is that the open magnet enables, as for the pinhole SPECT, an easy placement and a continuous visual monitoring of the small animal during the entire imaging procedure.
Sequential Versus Simultaneous SPECT/MRI
It is recognized that within the general field of molecular imaging there is a need for methods of coregistering MRI with nuclear imaging modalities, but use of high magnetic fields for simultaneous imaging (MRI and SPECT) places important constraints on many engineering and cost aspects of a system design as opposed to the dual device used in this study (3,10,31). The low-field MRI system and the MR-compatible animal holder we propose also had the advantages of being well adapted to an integrated coaxial rail-based dual device or a docking and click-over modular approach for combining sequential small-animal SPECT and MRI (32). Recent results have shown that semiconductor detectors (cadmium–zinc–telluride [CZT]) that can be used for small-animal SPECT were also insensitive to magnetic fields up to 7 T, leading to the development of a prototype limited to 24 unique angular samples but opening the way for simultaneous SPECT/MRI systems (3,10). Small-animal truly simultaneous versus sequential SPECT/MRI applications are actually difficult to define. This is mainly due to the constant technologic evolution for both pinhole SPECT and MR instrumentation to increase their sensitivities while minimizing the duration of image acquisition.
CONCLUSION
In this article we have shown that a small-animal SPECT/MRI dual hybrid device combining a pinhole SPECT imager adjacent to a small low-field MR imager is a low-cost potential solution for anatomic and functional combination that can be useful in molecular preclinical imaging.