FIGURE 4. Representative 511-keV energy spectrum acquired with LSO-APD detector in 7-T MRI scanner while applying spin-echo sequence. Energy resolution was 18.7% (511 keV, full width at half maximum). Although energy resolution is still good and linearity is maintained between the two 22Na peaks, MRI sequence causes increased noise in low-energy region.
|