View larger version (19K):
[in a new window]
|
FIGURE 2. Phantom with syringes filled with variable concentrations of gastrografin, plotted as measured emission data activity (in Bq/mL) versus measured contrast density by CT (in Hounsfield units [HU]). Maximum ( ) and mean ( ) measured activity with CT attenuation correction, and maximum ( ) and mean ( ) measured activity with 68Ge attenuation correction are displayed. Drop in measured activity is observed for high-density contrast. In this region, "real density" is >3,071 HU, but the PET/CT system plateaus at this level. Thus, overcorrection in presence of high-density contrast also has plateau corresponding to 3,071 HU density. However, because of increased true attenuation as result of increasing "real" density, measured counts decrease, resulting in drop in curve.
|