JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


First published online February 20, 2008, 10.2967/jnumed.107.043034
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
jnumed.107.043034v1
49/3/471    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in JNM
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tai, Y.-C.
Right arrow Articles by O'Sullivan, J. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tai, Y.-C.
Right arrow Articles by O'Sullivan, J. A.
Journal of Nuclear Medicine Vol. 49 No. 3 471-479
© 2008 by Society of Nuclear Medicine

doi: 10.2967/jnumed.107.043034

Basic Science Investigation

Virtual-Pinhole PET

Yuan-Chuan Tai1,2, Heyu Wu1, Debashish Pal3 and Joseph A. O'Sullivan4

1 Department of Radiology, Washington University, St. Louis, Missouri; 2 Alvin J. Siteman Cancer Center, Washington University School of Medicine, St. Louis, Missouri; 3 Department of Biomedical Engineering, Washington University, St. Louis, Missouri; and 4 Department of Electrical and System Engineering, Washington University, St. Louis, Missouri

Correspondence: For correspondence or reprints contact: Yuan-Chuan Tai, PhD, Washington University School of Medicine, Campus Box 8225, 510 S. Kingshighway Blvd., St. Louis, MO 63110. E-mail: taiy{at}wustl.edu

We proposed and tested a novel geometry for PET system design analogous to pinhole SPECT called the virtual-pinhole PET (VP-PET) geometry to determine whether it could provide high-resolution images. Methods: We analyzed the effects of photon acolinearity and detector sizes on system resolution and extended the empiric formula for reconstructed image resolution of conventional PET proposed earlier to predict the resolutions of VP-PET. To measure the system resolution of VP-PET, we recorded coincidence events as a 22Na point source was stepped across the coincidence line of response between 2 detectors made from identical arrays of 12 x 12 lutetium oxyorthosilicate crystals (each measuring 1.51 x 1.51 x 10 mm3) separated by 565 mm. To measure reconstructed image resolution, we built 4 VP-PET systems using 4 types of detectors (width, 1.51–6.4 mm) and imaged 4 point sources of 64Cu (half-life = 12.7 h to allow a long acquisition time). Tangential and radial resolutions were measured and averaged for each source and each system. We then imaged a polystyrene plastic phantom representing a 2.5-cm-thick cross-section of isolated breast volume. The phantom was filled with an aqueous solution of 64Cu (713 kBq/mL) in which the following were imbedded: 4 spheric tumors ranging from 1.8 to 12.6 mm in inner diameter (ID), 6 micropipettes (0.7- or 1.1-mm ID filled with 64Cu at 5x, 20x, or 50x background), and a 10.0-mm outer-diameter cold lesion. Results: The shape and measured full width at half maximum of the line spread functions agree well with the predicted values. Measured reconstructed image resolution (2.40–3.24 mm) was ±6% of the predicted value for 3 of the 4 systems. In one case, the difference was 12.6%, possibly due to underestimation of the block effect from the low-resolution detector. In phantom experiments, all spheric tumors were detected. Small line sources were detected if the activity concentration is at least 20x background. Conclusion: We have developed and characterized a novel geometry for PET. A PET system following the VP-PET geometry provides high-resolution images for objects near the system's high-resolution detectors. This geometry may lead to the development of special-purpose PET systems or resolution-enhancing insert devices for conventional PET scanners.

Key Words: PET • geometry • pinhole • breast imaging

COPYRIGHT © 2008 by the Society of Nuclear Medicine, Inc.


Related articles in JNM:

This Month in JNM

JNM 2008 49: 11A-12A. [Full Text]  

VP-PET: A New Imaging Modality?
Arion Chatziioannou
JNM 2008 49: 345-346. [Full Text]  



This article has been cited by other articles:


Home page
JNMHome page
H. Wu, D. Pal, T. Y. Song, J. A. O'Sullivan, and Y.-C. Tai
Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small-Animal PET Scanner
J. Nucl. Med., October 1, 2008; 49(10): 1668 - 1676.
[Abstract] [Full Text] [PDF]


Home page
JNMHome page
A. Chatziioannou
VP-PET: A New Imaging Modality?
J. Nucl. Med., March 1, 2008; 49(3): 345 - 346.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY THE JOURNAL OF NUCLEAR MEDICINE
Copyright © 2008 by the Society of Nuclear Medicine.