Brief Communications |
1 Division of Nuclear Medicine, Department of Internal Medicine, School of Medicine, St. Louis University, St. Louis, Missouri
2 Division of Nuclear Medicine, Department of Radiology, School of Medicine, The Johns Hopkins University, Baltimore, Maryland
ABSTRACT
This study evaluated lesion mislocalization between PET and CT on PET/CT studies when CT instead of germanium is used for attenuation correction (AC). Methods: PET/CT scans were obtained for 300 clinical patients. Both CT and germanium scans were used to correct PET emission data. Cases were noted of suspected inaccurate localization of lesions on any of the 5 sets of images (PET using germanium AC [GeAC] fused and not fused with CT, PET using CT AC fused and not fused with CT, and PET with no AC [NAC]). Independent CT or MRI was used to determine true lesion locations. Results: Six of 300 patients (2%) had lesion mislocalization when CT was used for AC or fusion. True liver dome lesions were mislocalized to the right lung base on PET/CT, likely because of a respiratory motion difference between PET and CT. No mislocalization was present on NAC PET or non-CT-fused GeAC PET images. Conclusion: Serious lesion mislocalization on PET/CT studies may occur, albeit very infrequently, when CT is used for either AC or fusion.
Key Words: PET/CT image fusion artifact respiratory motion
This article has been cited by other articles:
![]() |
D. W. Townsend Dual-Modality Imaging: Combining Anatomy and Function J. Nucl. Med., June 1, 2008; 49(6): 938 - 955. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. J. Pichler, H. F. Wehrl, and M. S. Judenhofer Latest Advances in Molecular Imaging Instrumentation J. Nucl. Med., June 1, 2008; 49(Suppl_2): 5S - 23S. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Suga, Y. Kawakami, H. Iwanaga, O. Tokuda, and N. Matsunaga Automated Breath-Hold Perfusion SPECT/CT Fusion Images of the Lungs Am. J. Roentgenol., August 1, 2007; 189(2): 455 - 463. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. V. Vogel, J. A. van Dalen, B. Wiering, H. Huisman, F. H.M. Corstens, T. J.M. Ruers, and W. J.G. Oyen Evaluation of Image Registration in PET/CT of the Liver and Recommendations for Optimized Imaging J. Nucl. Med., June 1, 2007; 48(6): 910 - 919. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S.P. Meirelles, Y. E. Erdi, S. A. Nehmeh, O. D. Squire, S. M. Larson, J. L. Humm, and H. Schoder Deep-Inspiration Breath-Hold PET/CT: Clinical Findings with a New Technique for Detection and Characterization of Thoracic Lesions J. Nucl. Med., May 1, 2007; 48(5): 712 - 719. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Martinez-Moller, M. Souvatzoglou, N. Navab, M. Schwaiger, and S. G. Nekolla Artifacts from Misaligned CT in Cardiac Perfusion PET/CT Studies: Frequency, Effects, and Potential Solutions J. Nucl. Med., February 1, 2007; 48(2): 188 - 193. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Blodgett, C. C. Meltzer, and D. W. Townsend PET/CT: Form and Function Radiology, February 1, 2007; 242(2): 360 - 385. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Brechtel and T. Beyer Reply: Adequate Evaluation of Image Registration in Hybrid PET/CT. J. Nucl. Med., September 1, 2006; 47(9): 1557 - 1558. [Full Text] [PDF] |
||||
![]() |
M. A. Blake, A. Singh, B. N. Setty, J. Slattery, M. Kalra, M. M. Maher, D. V. Sahani, A. J. Fischman, and P. R. Mueller Pearls and Pitfalls in Interpretation of Abdominal and Pelvic PET-CT RadioGraphics, September 1, 2006; 26(5): 1335 - 1353. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Delbeke, R. E. Coleman, M. J. Guiberteau, M. L. Brown, H. D. Royal, B. A. Siegel, D. W. Townsend, L. L. Berland, J. A. Parker, G. Zubal, et al. Procedure Guideline for SPECT/CT Imaging 1.0 J. Nucl. Med., July 1, 2006; 47(7): 1227 - 1234. [Full Text] [PDF] |
||||
![]() |
D. Delbeke, R. E. Coleman, M. J. Guiberteau, M. L. Brown, H. D. Royal, B. A. Siegel, D. W. Townsend, L. L. Berland, J. A. Parker, K. Hubner, et al. Procedure Guideline for Tumor Imaging with 18F-FDG PET/CT 1.0 J. Nucl. Med., May 1, 2006; 47(5): 885 - 895. [Full Text] [PDF] |
||||
![]() |
T. Pan, O. Mawlawi, S. A. Nehmeh, Y. E. Erdi, D. Luo, H. H. Liu, R. Castillo, R. Mohan, Z. Liao, and H.A. Macapinlac Attenuation Correction of PET Images with Respiration-Averaged CT Images in PET/CT J. Nucl. Med., September 1, 2005; 46(9): 1481 - 1487. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Coleman, D. Delbeke, M. J. Guiberteau, P. S. Conti, H. D. Royal, J. C. Weinreb, B. A. Siegel, M. F. Federle, D. W. Townsend, and L. L. Berland Concurrent PET/CT with an Integrated Imaging System: Intersociety Dialogue from the Joint Working Group of the American College of Radiology, the Society of Nuclear Medicine, and the Society of Computed Body Tomography and Magnetic Resonance J. Nucl. Med., July 1, 2005; 46(7): 1225 - 1239. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. F. Branstetter IV, T. M. Blodgett, L. A. Zimmer, C. H. Snyderman, J. T. Johnson, S. Raman, and C. C. Meltzer Head and Neck Malignancy: Is PET/CT More Accurate than PET or CT Alone? Radiology, May 1, 2005; 235(2): 580 - 586. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Blodgett, B. Casagranda, D. W. Townsend, and C. C. Meltzer Issues, Controversies, and Clinical Utility of Combined PET/CT Imaging: What Is the Interpreting Physician Facing? Am. J. Roentgenol., May 1, 2005; 184(5_supp): S138 - S145. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Boucher, S. Rodrigue, R. Lecomte, and F. Benard Respiratory Gating for 3-Dimensional PET of the Thorax: Feasibility and Initial Results J. Nucl. Med., February 1, 2004; 45(2): 214 - 219. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. T. Thomas, P. T. Bradshaw, B. H. Pollock, J. E. Montie, J. M.G. Taylor, H. D. Thames, P. W. McLaughlin, D. A. DeBiose, D. H. Hussey, and R. L. Wahl In Reply: J. Clin. Oncol., January 15, 2004; 22(2): 380 - 381. [Full Text] [PDF] |
||||
![]() |
R. S. Hattner Ahem, Blame the Daughter! J. Nucl. Med., January 1, 2004; 45(1): 153 - 153. [Full Text] [PDF] |
||||
![]() |
D. W. Townsend, J. P.J. Carney, J. T. Yap, and N. C. Hall PET/CT Today and Tomorrow J. Nucl. Med., January 1, 2004; 45(90010): 4S - 14. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. V. Vogel, W. J.G. Oyen, J. O. Barentsz, J. H.A.M. Kaanders, and F. H.M. Corstens PET/CT: Panacea, Redundancy, or Something in Between? J. Nucl. Med., January 1, 2004; 45(90010): 15S - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Wahl Why Nearly All PET of Abdominal and Pelvic Cancers Will Be Performed as PET/CT J. Nucl. Med., January 1, 2004; 45(90010): 82S - 95. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Antoch, F. M. Vogt, L. S. Freudenberg, F. Nazaradeh, S. C. Goehde, J. Barkhausen, G. Dahmen, A. Bockisch, J. F. Debatin, and S. G. Ruehm Whole-Body Dual-Modality PET/CT and Whole-Body MRI for Tumor Staging in Oncology JAMA, December 24, 2003; 290(24): 3199 - 3206. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |