JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH RSS TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sankaran, S.
Right arrow Articles by Tsui, B. M.W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sankaran, S.
Right arrow Articles by Tsui, B. M.W.
Journal of Nuclear Medicine Vol. 43 No. 3 432-438
© 2002 by Society of Nuclear Medicine


Basic Science Investigations

Optimum Compensation Method and Filter Cutoff Frequency in Myocardial SPECT: A Human Observer Study

Sharlini Sankaran, MS;1, Eric C. Frey, PhD;1, Karen L. Gilland, PhD;1 and Benjamin M.W. Tsui, PhD1

1 Department of Biomedical Engineering, School of Medicine, The University of North Carolina at Chapel Hill, Chapel Hill, North Carolina

Attenuation, photon scatter, and distance-dependent collimator-detector response are major degrading factors in myocardial SPECT images. The current study investigated whether compensation for these factors improves perfusion defect detectability, and compared the results for human observers with a previous study using a mathematical observer. Methods: Four methods were investigated: attenuation compensation (AC); attenuation and detector response compensation; attenuation and scatter compensation; and attenuation, detector response, and scatter compensation (ADSC). For ADSC, 4 three-dimensional postreconstruction Butterworth filter cutoff frequencies were investigated for a pixel size of 0.62 cm: 0.12, 0.14, 0.16, and 0.22 pixel-1. Five observers read images reconstructed using the 4 compensation methods. Receiver operating characteristics (ROC) analysis was used to determine the area under the ROC curve in each treatment studied. Results: Reconstruction methods that incorporated scatter and detector response compensation had higher indices of detectability than AC alone. Over the range studied, a filter cutoff frequency of 0.14 pixel-1 was optimal. A comparison of human observer results with an earlier channelized Hotelling observer study performed with the same images showed excellent agreement in trend and ranking of defect detectability. Conclusion: Compensation for detector response and scatter improves defect detectability compared with AC alone, although detectability may depend on phantom population choice and noise level. An optimal filter cutoff was found that is lower than what is typically used in a clinical setting. The channelized Hotelling observer is a good predictor of human observer performance and may reduce the need for tedious, time-consuming studies with human observers.

Key Words: myocardial SPECT • observer study • receiver operating characteristics • channelized Hotelling observer




This article has been cited by other articles:


Home page
J. Nucl. Med. Technol.Home page
Y. Yan and G. L. Zeng
A Postprocessing Method for Compensation of Scatter and Collimator Blurring in SPECT: A Proof-of-Concept Study
J. Nucl. Med. Technol., June 1, 2009; 37(2): 83 - 90.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH RSS TABLE OF CONTENTS
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY THE JOURNAL OF NUCLEAR MEDICINE
Copyright © 2002 by the Society of Nuclear Medicine.