JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH RSS TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Matsunari, I.
Right arrow Articles by Hisada, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Matsunari, I.
Right arrow Articles by Hisada, K.
Journal of Nuclear Medicine Vol. 42 No. 10 1579-1585
© 2001 by Society of Nuclear Medicine


Basic Science Investigations

Phantom Studies for Estimation of Defect Size on Cardiac 18F SPECT and PET: Implications for Myocardial Viability Assessment

Ichiro Matsunari, Tatsuya Yoneyama, Sugako Kanayama, Masamichi Matsudaira, Kenichi Nakajima, Junichi Taki, Stephan G. Nekolla, Norihisa Tonami and Kinichi Hisada

The Medical and Pharmacological Research Center Foundation, Ishikawa; Department of Nuclear Medicine, Kanazawa University School of Medicine, Kanazawa; Department of Cardiology, Kanazawa Medical University, Ishikawa, Japan; and Nuklearmedizinische Klinik und Poliklinik Technische Universität München, Munich, Germany

SPECT with 18F-FDG has emerged as an alternative to dedicated PET for the assessment of myocardial viability. However, whether FDG SPECT can reliably quantify the extent of viable and scarred myocardium is uncertain. The aim of this study was to investigate whether SPECT with an 18F-labeled agent would provide information on defect size similar to that provided by dedicated PET. Methods: Imaging was performed using an elliptic cylinder chest phantom with simulated bone, lung, mediastinum, liver, and heart. 18F was administered into the myocardium, mediastinum, right and left ventricular cavities, and liver. Plastic inserts (n = 11) ranging in size from 2% to 60% of the myocardium were used to simulate transmural myocardial infarctions. The chest phantom was imaged with a dedicated PET camera and with a double-head SPECT camera equipped with ultra-high-energy collimators. Both SPECT and PET data were analyzed using a semiquantitative polar map approach. Defects were quantified using various cutoff thresholds ranging from 30% to 80% of peak activity and were expressed as a percentage of the left ventricular myocardium. Defect size as measured by SPECT or PET was compared with true defect size. Results: The measured SPECT defect size was highly variable depending on the cutoff used, whereas PET defect size was relatively constant over the range of cutoffs tested. The mean absolute difference between measured and true defect sizes was minimal at a cutoff of 50% of peak activity for both SPECT (3.3% ± 3.3%) and PET (2.7% ± 2.5%). For this threshold, both SPECT and PET measurements showed an excellent correlation with true defect size (r = 0.98 for SPECT and 0.99 for PET). The correlation between SPECT and PET measurements was also excellent (r = 0.99; P < 0.01). Conclusion: If an appropriate threshold is used to define a defect, SPECT with an 18F-labeled agent can accurately measure defect size similarly to the manner of PET.

Key Words: SPECT • PET • 18F-FDG




This article has been cited by other articles:


Home page
The Computer JournalHome page
H. Greenspan
Super-Resolution in Medical Imaging
The Computer Journal, January 1, 2009; 52(1): 43 - 63.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH RSS TABLE OF CONTENTS
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY THE JOURNAL OF NUCLEAR MEDICINE
Copyright © 2001 by the Society of Nuclear Medicine.