JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


The Journal of Nuclear Medicine Vol. 40 No. 5 783-785
© 1999 by Society of Nuclear Medicine
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Van den Eeckhaut, A. C.
Right arrow Articles by Dierckx, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Van den Eeckhaut, A. C.
Right arrow Articles by Dierckx, R. A.

Laparoscopie Adjustable Silicone Gastric Banding Leakage Assessed by 99mTc-Pertechnetate Scintigraphy

Anja C. Van den Eeckhaut, Geert M. Villeirs, Christophe Van de Wiele, Jan J. Versijpt, Piet Pattyn and Rudi A. Dierckx

Division of Nuclear Medicine, Departments of Radiology and Surgery, University Hospital Gent, Gent, Belgium

Correspondence: For correspondence or reprints contact: Anja C. Van den Eeckhaut, MD, Division of Nuclear Medicine, University Hospital Gent, De Pintelaan 185, B-9000 Gent, Belgium.

ABSTRACT

The purpose of this study was to assess the value of 99mTc-pertechnetate scanning in the diagnosis of gastric banding leaks. Methods: Three patients with morbid obesity received laparoscopic adjustable silicone gastric banding (ASGB), but no significant weight reduction was obtained. To exclude band leakage as the cause, four scintigraphic procedures were performed, consisting of imaging the upper abdomen 30 min and 3 h after injection of 3 mL (111 MBq) pertechnetate solution into the ASGB reservoir. In one patient, the integrity of the ASGB device was first assessed radiologically after injection of a water-soluble contrast agent into the ASGB reservoir. Results: In two normally function ing ASGB devices, radiotracer was observed within the device on both early and late images. In two patients with a surgically proven small leak in the reservoir or the connecting tube, late images showed little tracer in the reservoir and the connecting tube. However, intense tracer accumulation was observed in the stomach as a result of resorption of pertechnetate in the subcutaneous or peritoneal blood vessels and subsequent gastric uptake. In one of the latter patients, radiographic assessment of the ASGB device revealed no abnormalities. Conclusion: 99mTc-pertechnetate scanning is a valuable technique to diagnose small leaks in an ASGB device.

Key Words: gastric banding • complications • obesity • 99mTc-pertechnetate scanning




This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
M. J. Mehanna, G. Birjawi, H. A. Moukaddam, G. Khoury, M. Hussein, and A. Al-Kutoubi
Complications of Adjustable Gastric Banding, a Radiological Pictorial Review