JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


The Journal of Nuclear Medicine Vol. 40 No. 4 604-613
© 1999 by Society of Nuclear Medicine
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Case, J. A.
Right arrow Articles by Weaver, J. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Case, J. A.
Right arrow Articles by Weaver, J. P.

Bone SPECT of the Spine: A Comparison of Attenuation Correction Techniques

James A. Case, Robert Licho, Michael A. King and John P. Weaver

Departments of Nuclear Medicine and Surgery, University of Massachusetts Medical Center, Worcester, Massachusetts

Correspondence: For correspondence or reprints contact: Robert Licho, MD, Department of Nuclear Medicine, University of Massachusetts Medical Center, 55 Lake Ave. North, Worcester, MA 01655.

ABSTRACT

Image artifacts from variable self-attenuation are recognized as major sources of diagnostic uncertainty in SPECT. For myocardial perfusion studies, an attenuation map is often obtained from a separate transmission study. However, for many applications such as bone SPECT, it has been believed to be unnecessary to obtain a transmission study to correct for the effects of attenuation. We have had significant success in clinical management of lower spine pain using bone SPECT. This success has led us to consider SPECT for the management of cervical spine pain. Cervical spine reconstructions without attenuation correction are difficult to interpret, because the high attenuation in the mandible and skull tends to decrease estimates of activity of the upper cervical spine, and the lower cervical/upper thoracic vertebrae are obscured by the shoulders. We present a technique that uses downscatter to provide attenuation correction for these acquisitions and compare it with other recognized attenuation correction techniques. Methods: An emission study is acquired using two windows: one for obtaining the photopeak data and another for obtaining the downscattered photons. A body outline is estimated from these datasets using a projection data thresholding method. From this outline, a uniform attenuation map is created using attenuation coefficients appropriate for 99mTc in water (0.154 cm–1). These maps are used in SPECT reconstruction using ordered-subset expectation maximization (OSEM). This method is compared with (a) no attenuation correction (NC), (b) conventional Chang attenuation correction based on the interactive determination of the body outline from the 99mTc emission photopeak data (ChangAC) and (c) OSEM correction using attenuation maps estimated with a line source and fanbeam collimators (transAC). Results: Patient studies using scatterAC demonstrated a significant improvement in the uniformity of estimated cervical spine uptake in normal patients, compared with either NC or ChangAC. Results using scatterAC were similar to those of transAC. We also observed significant improvement in uniformity using scatterAC in SPECT of the lower back in obese patients, as well as the relative limitations of scatterAC versus nonuniform, transmission-based attenuation correction. Conclusion: Comparisons with reconstructions using transmission data for estimating attenuation demonstrate that reasonable quantitative accuracy can be obtained in SPECT of the cervical spine using this simple attenuation estimate. Both scatterAC and transAC appear to provide consistent and expected uniform spine uptake in the cervical spines of normal patients.

Key Words: bone SPECT • spine • attenuation • back pain




This article has been cited by other articles:


Home page
JNMHome page
R. G. Wells, T. Farncombe, E. Chang, and R. L. Nicholson