Research Program Radiological Diagnostics and Therapy, German Cancer Research Center (DKFZ), Heidelberg, Germany
Correspondence: For correspondence contact: Gunnar Brix, MD, Forschungsschwerpunkt "Radiologische Diagnostik und Therapie," Deutsches Krebsforschungszentrum, Im Neuenheimer Feld 280, D-69120 Heidelbergm, Germany.
ABSTRACT
This study evaluates the performance of the newly developed high-resolution whole-body PET scanner ECAT EXACT HR+. Methods: The scanner consists of four rings of 72 bismuth germanate block detectors each, covering an axial field of view of 15.5 cm with a patient port of 56.2 cm. A single block detector is divided into an 8 x 8 matrix, giving a total of 32 rings with 576 detectors each. The dimensions of a single detector element are 4.39 x 4.05 x 30 mm3. The scanner is equipped with extendable tungsten septa for two-dimensional two-dimensional measurements, as well as with three 68Ge line sources for transmission scans and daily quality control. The spatial resolution, scatter fraction, count rate, sensitivity, uniformity and accuracy of the implemented correction algorithms were evaluated after the National Electrical Manufacturers Association protocol using the standard acquisition parameters. Results: The transaxial resolution in the two-dimensional mode is 4.3 mm (4.4 mm) in the center and increases to 4.7 mm (4.8 mm) tangential and to 8.3 mm (8.0 mm) radial at a distance of r = 20 cm from the center. The axial slice width measured in the two-dimensional mode varies between 4.2 and 6.6 mm FWHM over the transaxial field of view. In the three-dimensional mode the average axial resolution varies between 4.1 mm FWHM in the center and 7.8 mm at r = 20 cm. The scatter fraction is 17.1% (32.5%) for a lower energy discriminator level of 350 keV. The maximum true event count rate of 263 (345) kcps was measured at an activity concentration of 142 (26.9) kBq/ml. The total system sensitivity for true events is 5.7 (27.7) cps/Bq/ml. From the uniformity measurements, we obtained a volume variance of 3.9% (5.0%) and a system variance of 1.6% (1.7%). The implemented three-dimensional scatter correction algorithm reveals very favorable properties, whereas the three-dimensional attenuation correction yields slightly inaccurate results in low-and high-density regions. Conclusion: The ECAT EXACT HR+ has an excellent, nearly isotropie spatial resolution, which is advantageous for brain and small animal studies. While the relatively low slice sensitivity may hamper the capability for performing fast dynamic two-dimensional studies, the scanner offers a sufficient sensitivity and count rate capacity for fully three-dimensional whole-body imaging.
Key Words: PET NEMA Protocol
This article has been cited by other articles:
![]() |
V. Bouilleret, F. Semah, F. Chassoux, M. Mantzaridez, A. Biraben, R. Trebossen, and M. -J. Ribeiro Basal ganglia involvement in temporal lobe epilepsy: A functional and morphologic study Neurology, January 15, 2008; 70(3): 177 - 184. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schepis, O. Gaemperli, V. Treyer, I. Valenta, C. Burger, P. Koepfli, M. Namdar, I. Adachi, H. Alkadhi, and P. A. Kaufmann Absolute Quantification of Myocardial Blood Flow with 13N-Ammonia and 3-Dimensional PET J. Nucl. Med., November 1, 2007; 48(11): 1783 - 1789. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Edison, H. A. Archer, R. Hinz, A. Hammers, N. Pavese, Y. F. Tai, G. Hotton, D. Cutler, N. Fox, A. Kennedy, et al. Amyloid, hypometabolism, and cognition in Alzheimer disease: An [11C]PIB and [18F]FDG PET study Neurology, February 13, 2007; 68(7): 501 - 508. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. M. Scheinin, T. K. Tolvanen, I. A. Wilson, E. M. Arponen, K. A. Nagren, and J. O. Rinne Biodistribution and Radiation Dosimetry of the Amyloid Imaging Agent 11C-PIB in Humans J. Nucl. Med., January 1, 2007; 48(1): 128 - 133. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. J. Kemp, C. Kim, J. J. Williams, A. Ganin, and V. J. Lowe NEMA NU 2-2001 Performance Measurements of an LYSO-Based PET/CT System in 2D and 3D Acquisition Modes J. Nucl. Med., December 1, 2006; 47(12): 1960 - 1967. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lubberink, A. van Schie, H. W.A.M. de Jong, G. A.M.S. van Dongen, and G. J.J. Teule Acquisition Settings for PET of 124I Administered Simultaneously with Therapeutic Amounts of 131I J. Nucl. Med., August 1, 2006; 47(8): 1375 - 1381. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Luaute, C. Michel, G. Rode, L. Pisella, S. Jacquin-Courtois, N. Costes, F. Cotton, D. le Bars, D. Boisson, P. Halligan, et al. Functional anatomy of the therapeutic effects of prism adaptation on left neglect Neurology, June 27, 2006; 66(12): 1859 - 1867. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Picchio, U. Treiber, A. J. Beer, S. Metz, P. Bossner, H. van Randenborgh, R. Paul, G. Weirich, M. Souvatzoglou, R. Hartung, et al. Value of 11C-Choline PET and Contrast-Enhanced CT for Staging of Bladder Cancer: Correlation with Histopathologic Findings J. Nucl. Med., June 1, 2006; 47(6): 938 - 944. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Meisetschlager, T. Poethko, A. Stahl, I. Wolf, K. Scheidhauer, M. Schottelius, M. Herz, H. J. Wester, and M. Schwaiger Gluc-Lys([18F]FP)-TOCA PET in Patients with SSTR-Positive Tumors: Biodistribution and Diagnostic Evaluation Compared with [111In]DTPA-Octreotide J. Nucl. Med., April 1, 2006; 47(4): 566 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Lodge, R. D. Badawi, R. Gilbert, P. E. Dibos, and B. R. Line Comparison of 2-Dimensional and 3-Dimensional Acquisition for 18F-FDG PET Oncology Studies Performed on an LSO-Based Scanner J. Nucl. Med., January 1, 2006; 47(1): 23 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Matsumoto, K. Kitamura, T. Mizuta, K. Tanaka, S. Yamamoto, S. Sakamoto, Y. Nakamoto, M. Amano, K. Murase, and M. Senda Performance Characteristics of a New 3-Dimensional Continuous-Emission and Spiral-Transmission High-Sensitivity and High-Resolution PET Camera Evaluated with the NEMA NU 2-2001 Standard J. Nucl. Med., January 1, 2006; 47(1): 83 - 90. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schwaiger, S. Ziegler, and S. G. Nekolla PET/CT: Challenge for Nuclear Cardiology J. Nucl. Med., October 1, 2005; 46(10): 1664 - 1678. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-C. Richard, M. Janier, F. Lavenne, C. Tourvieille, D. Le Bars, N. Costes, G. Gimenez, and C. Guerin Quantitative Assessment of Regional Alveolar Ventilation and Gas Volume Using 13N-N2 Washout and PET J. Nucl. Med., August 1, 2005; 46(8): 1375 - 1383. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bouvard, N. Costes, F. Bonnefoi, F. Lavenne, F. Mauguiere, J. Delforge, and P. Ryvlin Seizure-related short-term plasticity of benzodiazepine receptors in partial epilepsy: a [11C]flumazenil-PET study Brain, June 1, 2005; 128(6): 1330 - 1343. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Lehrer, B. T. Christian, J. Mantil, A. C. Murray, B. R. Buchsbaum, T. R. Oakes, W. Byne, E. M. Kemether, and M. S. Buchsbaum Thalamic and Prefrontal FDG Uptake in Never Medicated Patients With Schizophrenia Am J Psychiatry, May 1, 2005; 162(5): 931 - 938. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Henze, A. Dimitrakopoulou-Strauss, S. Milker-Zabel, J. Schuhmacher, L. G. Strauss, J. Doll, H. R. Macke, M. Eisenhut, J. Debus, and U. Haberkorn Characterization of 68Ga-DOTA-D-Phe1-Tyr3-Octreotide Kinetics in Patients with Meningiomas J. Nucl. Med., May 1, 2005; 46(5): 763 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. A. Mullani and O. Mawlawi Why Is the Resolution of the Discovery PET/CT Camera So Poor? J. Nucl. Med., May 1, 2005; 46(5): 898 - 898. [Full Text] [PDF] |
||||
![]() |
T. Beyer, L. Tellmann, I. Nickel, and U. Pietrzyk On the Use of Positioning Aids to Reduce Misregistration in the Head and Neck in Whole-Body PET/CT Studies J. Nucl. Med., April 1, 2005; 46(4): 596 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Bouilleret, F. Semah, A. Biraben, D. Taussig, F. Chassoux, A. Syrota, and M.-J. Ribeiro Involvement of the Basal Ganglia in Refractory Epilepsy: An 18F-Fluoro-L-DOPA PET Study Using 2 Methods of Analysis J. Nucl. Med., March 1, 2005; 46(3): 540 - 547. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-D. Gallezot, M. Bottlaender, M.-C. Gregoire, D. Roumenov, J.-R. Deverre, C. Coulon, M. Ottaviani, F. Dolle, A. Syrota, and H. Valette In Vivo Imaging of Human Cerebral Nicotinic Acetylcholine Receptors with 2-18F-Fluoro-A-85380 and PET J. Nucl. Med., February 1, 2005; 46(2): 240 - 247. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lubberink, R. Boellaard, A. P. van der Weerdt, F. C. Visser, and A. A. Lammertsma Quantitative Comparison of Analytic and Iterative Reconstruction Methods in 2- and 3-Dimensional Dynamic Cardiac 18F-FDG PET J. Nucl. Med., December 1, 2004; 45(12): 2008 - 2015. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Mazzadi, M. F. Janier, B. Brossier, X. Andre-Fouet, D. Revel, and P. Croisille Tagged MRI and PET in severe CAD: discrepancy between preoperative inotropic reserve and intramyocardial functional outcome after revascularization Am J Physiol Heart Circ Physiol, November 1, 2004; 287(5): H2226 - H2233. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. De Tiege, S. Goldman, S. Laureys, D. Verheulpen, C. Chiron, C. Wetzburger, P. Paquier, D. Chaigne, N. Poznanski, I. Jambaque, et al. Regional cerebral glucose metabolism in epilepsies with continuous spikes and waves during sleep Neurology, September 14, 2004; 63(5): 853 - 857. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Boellaard, N. C. Krak, O. S. Hoekstra, and A. A. Lammertsma Effects of Noise, Image Resolution, and ROI Definition on the Accuracy of Standard Uptake Values: A Simulation Study J. Nucl. Med., September 1, 2004; 45(9): 1519 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Rohren, T. G. Turkington, and R. E. Coleman Clinical Applications of PET in Oncology Radiology, May 1, 2004; 231(2): 305 - 332. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Henze, A. Mohammed, H. P. Schlemmer, K. K. Herfarth, S. Hoffner, S. Haufe, W. Mier, M. Eisenhut, J. Debus, and U. Haberkorn PET and SPECT for Detection of Tumor Progression in Irradiated Low-Grade Astrocytoma: A Receiver-Operating-Characteristic Analysis J. Nucl. Med., April 1, 2004; 45(4): 579 - 586. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. P. van der Weerdt, R. Boellaard, P. Knaapen, C. A. Visser, A. A. Lammertsma, and F. C. Visser Postinjection Transmission Scanning in Myocardial 18F-FDG PET Studies Using Both Filtered Backprojection and Iterative Reconstruction J. Nucl. Med., February 1, 2004; 45(2): 169 - 175. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bai, P. E. Kinahan, D. Brasse, C. Comtat, D. W. Townsend, C. C. Meltzer, V. Villemagne, M. Charron, and M. Defrise An Analytic Study of the Effects of Attenuation on Tumor Detection in Whole-Body PET Oncology Imaging J. Nucl. Med., November 1, 2003; 44(11): 1855 - 1861. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Tarantola, F. Zito, and P. Gerundini PET Instrumentation and Reconstruction Algorithms in Whole-Body Applications J. Nucl. Med., May 1, 2003; 44(5): 756 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lartizien, P. E. Kinahan, R. Swensson, C. Comtat, M. Lin, V. Villemagne, and R. Trebossen Evaluating Image Reconstruction Methods for Tumor Detection in 3-Dimensional Whole-Body PET Oncology Imaging J. Nucl. Med., February 1, 2003; 44(2): 276 - 290. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Fias, J. Lammertyn, B. Reynvoet, P. Dupont, and G. A. Orban Parietal Representation of Symbolic and Nonsymbolic Magnitude J. Cogn. Neurosci., January 1, 2003; 15(1): 47 - 56. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Richard, M. Janier, F. Lavenne, V. Berthier, D. Lebars, G. Annat, F. Decailliot, and C. Guerin Effect of position, nitric oxide, and almitrine on lung perfusion in a porcine model of acute lung injury J Appl Physiol, December 1, 2002; 93(6): 2181 - 2191. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kadrmas and P. E. Christian Comparative Evaluation of Lesion Detectability for 6 PET Imaging Platforms Using a Highly Reproducible Whole-Body Phantom with 22Na Lesions and Localization ROC Analysis J. Nucl. Med., November 1, 2002; 43(11): 1545 - 1554. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lartizien, C. Comtat, P. E. Kinahan, N. Ferreira, B. Bendriem, and R. Trebossen Optimization of Injected Dose Based on Noise Equivalent Count Rates for 2- and 3-Dimensional Whole-Body PET J. Nucl. Med., September 1, 2002; 43(9): 1268 - 1278. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. P. Schafers, T. J. Spinks, P. G. Camici, P. M. Bloomfield, C. G. Rhodes, M. P. Law, C. S.R. Baker, and O. Rimoldi Absolute Quantification of Myocardial Blood Flow with H215O and 3-Dimensional PET: An Experimental Validation J. Nucl. Med., August 1, 2002; 43(8): 1031 - 1040. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-C. Richard, M. Janier, F. Decailliot, D. Le Bars, F. Lavenne, V. Berthier, M. Lionnet, L. Cinotti, G. Annat, and C. Guerin Comparison of PET with Radioactive Microspheres to Assess Pulmonary Blood Flow J. Nucl. Med., August 1, 2002; 43(8): 1063 - 1071. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Fias, P. Dupont, B. Reynvoet, and G. A. Orban The Quantitative Nature of a Visual Task Differentiates between Ventral and Dorsal Stream J. Cogn. Neurosci., May 1, 2002; 14(4): 646 - 658. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Nofzinger, D. J. Buysse, J. M. Miewald, C. C. Meltzer, J. C. Price, R. C. Sembrat, H. Ombao, C. F. Reynolds, T. H. Monk, M. Hall, et al. Human regional cerebral glucose metabolism during non-rapid eye movement sleep in relation to waking Brain, May 1, 2002; 125(5): 1105 - 1115. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Pieterman, T. H. Que, P. H. Elsinga, J. Pruim, J. W.G. van Putten, A. T.M. Willemsen, W. Vaalburg, and H. J.M. Groen Comparison of 11C-Choline and 18F-FDG PET in Primary Diagnosis and Staging of Patients with Thoracic Cancer J. Nucl. Med., February 1, 2002; 43(2): 167 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Cornette, P. Dupont, E. Salmon, and G. A. Orban The Neural Substrate of Orientation Working Memory J. Cogn. Neurosci., August 1, 2001; 13(6): 813 - 828. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Boellaard, A. van Lingen, and A. A. Lammertsma Experimental and Clinical Evaluation of Iterative Reconstruction (OSEM) in Dynamic PET: Quantitative Characteristics and Effects on Kinetic Modeling J. Nucl. Med., May 1, 2001; 42(5): 808 - 817. [Abstract] [Full Text] |
||||
![]() |
T. G. Turkington Introduction to PET Instrumentation J. Nucl. Med. Technol., March 1, 2001; 29(1): 4 - 11. [Abstract] [Full Text] |
||||
![]() |
L. Cornette, P. Dupont, G. Bormans, L. Mortelmans, and G.A. Orban Separate Neural Correlates for the Mnemonic Components of Successive Discrimination and Working Memory Tasks Cereb Cortex, January 1, 2001; 11(1): 59 - 72. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. C. Meltzer Neuropharmacology and Receptor Studies in the Elderly J Geriatr Psychiatry Neurol, October 1, 1999; 12(3): 137 - 149. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |