U66 INSERM, Institut Gustave-Roussy, Villejuif, France
Department of Medical Physics, University College London, London, United Kingdom
Correspondence: For correspondence or reprints contact: Irène Buvat, U66 INSERM, Institut Gustave-Roussy, 39 rue Camille Desmoulins, 94 805 Villejuif Cedex, France.
ABSTRACT
We compared nine scatter correction methods based on spectral analysis which process SPECT projections. Methods: Monte Carlo simulation was used to generate histories of photons emitted from a realistic 99mTc phantom. A particular projection was considered. Information regarding the history, location and energy of the photons detected in this projection was analyzed to test the assumptions underlying each scatter correction method. Relative and absolute quantification and signal-to-noise ratio were assessed for each scatter corrected image. Results: For the simulated data, two methods do not enable activity quantification. Among the methods requiring some parameters to be calibrated, the dual-energy window method shows the best compromise between accuracy and ease of implementation but introduces a bias in relative quantification. In this respect, a triple-energy window technique is more accurate than the dual-window method. A factor analysis approach results in more stable quantitative accuracy (error
10%) for a wide range of activity but requires a more sophisticated acquisition mode (30 energy windows). Conclusion: These results show that a scatter correction method using spectral analysis can be used to substantially improve accurate quantification.
Key Words: scatter correction single-photon emission computed tomography spectral analysis
This article has been cited by other articles:
![]() |
I. Aubineau-Laniece, L. de Carlan, I. Clairand, A. Lemosquet, S. Chiavassa, N. Pierrat, M. Bardies, and D. Franck Current developments at IRSN on computational tools dedicated to assessing doses for both internal and external exposure Radiat Prot Dosimetry, December 20, 2005; 115(1-4): 522 - 529. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Cot, C. Falcon, C. Crespo, J. Sempau, D. Pareto, S. Bullich, F. Lomena, F. Calvino, J. Pavia, and D. Ros Absolute Quantification in Dopaminergic Neurotransmission SPECT Using a Monte Carlo-Based Scatter Correction and Fully 3-Dimensional Reconstruction J. Nucl. Med., September 1, 2005; 46(9): 1497 - 1504. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Vines, M. Ichise, J.-S. Liow, H. Toyama, and R. B. Innis Evaluation of 2 Scatter Correction Methods Using a Striatal Phantom for Quantitative Brain SPECT J. Nucl. Med. Technol., September 1, 2003; 31(3): 157 - 160. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |