JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


The Journal of Nuclear Medicine Vol. 33 No. 5 735-738
© 1992 by Society of Nuclear Medicine
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Roberson, P.L.
Right arrow Articles by Ensminger, W.D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Roberson, P.L.
Right arrow Articles by Ensminger, W.D.

Three-Dimensional Tumor Dosimetry for Hepatic Yttrium-90-Microsphere Therapy

P.L. Roberson, R.K. Ten Haken, D.L. McShan, P.E. McKeever and W.D. Ensminger

Departments of Radiation Oncology, Pathology and Internal Medicine, University of Michigan Medical Center, Ann Arbor, Michigan

Correspondence: For reprints contact: Peter L. Roberson, PhD, Department of Radiation Oncology. University of Michigan Medical School, UH-B2C490, Box 0010, 1500 E. Medical Center Dr., Ann Arbor, MI 48109.

ABSTRACT

The nonuniformity of dose deposition for hepatuc 90Y-microsphere therapy is believed to play an important role in the relative sparing of normal liver tissues. To help study this issue, three-dimensional dose calculations have been performed for the VX2 tumor model in the rabbit treated with hepatic arterial administration of 90Y-glass microspheres (90Y-MS). Colored, nonactivated spheres of similar size to 90Y-MS were injected into the hepatic artery to mimic the treatment deposition of 90Y-MS. Sample blocks of treated liver were serially sectioned (200 µm thickness), fixed and photographed showing the position of the colored microspheres. The microsphere positions were digitized into a three-dimensional treatment planning system, and three-dimensional dose calculations were performed. A 2-mm diameter liver tumor nodule receiving 15 times more microspheres than nearby normal liver resulted in tumor-to-normal-tissue (TNT) calculated dose ratios of 2.6 (average dose) and 1.9 (minimum dose). The nonuniform microsphere distribution resulted in a dose gradient over the nodule with a minimum value which was less than one half the average dose. The relative dose deposition in the vicinity of the tumor nodule does not fully reconcile the known liver tolerance dose derived from uniform irradiation with the large calculated average doses tolerated with this type of therapy.




This article has been cited by other articles:


Home page
JNMHome page
J. E. Dancey, F. A. Shepherd, K. Paul, K. W. Sniderman, S. Houle, J. Gabrys, A. L. Hendler, and J. E. Goin
Treatment of Nonresectable Hepatocellular Carcinoma with Intrahepatic 90Y-Microspheres
J. Nucl. Med., October 1, 2000; 41(10): 1673 - 1681.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY THE JOURNAL OF NUCLEAR MEDICINE
Copyright © 1992 by the Society of Nuclear Medicine.