Nuclear Medicine Department, William Beaumont Hospital, Royal Oak, Michigan
Correspondence: For reprints contact: John E. Freitas, MD, Nuclear Medicine Dept., William Beaumont Hospital, 3601 West 13 Mile Rd., Royal Oak, MI 48072.
ABSTRACT
Malfunction of a computer-assisted digital gamma camera is reported. Despite what appeared to be adequate acceptance testing, an error in the system gave rise to switching of images and identification text. A suggestion is made for using a hot marker, which would avoid the potential error of misinterpretation of patient images.