Department of Radiology, University of Southern California, Los Angeles, California
Correspondence: For reprints contact: Manbir Singh, PhD, Dept. of Radiology, University of Southern California, PSC-610, 1985 Zonal Ave., Los Angeles, CA-90033.
ABSTRACT
A collimated germanium detector with an energy resolution of 1 keV full width at half maximum at 140 keV was used to measure the energy spectrum of radiation emitted from a test object containing an asymmetric distribution of 99mTc and nonuniform attenuation. Energy spectra were recorded from 24 positions around the object and convolved with Gaussian functions to simulate data that would have been acquired with a scintillation camera. The scatter fraction was computed from the convolved spectra in conjunction with a scatter-free reference spectrum. After adding appropriate Poisson noise, a technique based on maximizing the signal to noise ratio was developed to optimally subtract the scatter fraction from the recorded counts. SPECT imaging of the test object was performed to evaluate the correction technique.
FOOTNOTES
A portion of this paper was presented at the SNM 33rd Annual meeting, June 2225, 1986, Washington DC.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |