Albert Einstein College of Medicine, Bronx, New York
Correspondence: For reprints contact: Sherman L. Heller, Ph.D., Albert Einstein College of Medicine, Dept. of Nuclear Medicine, 1300 Morris Park Avenue, Bronx, NY 10461.
ABSTRACT
A dual collimator was designed for an Anger camera to permit measurement of cardiac performance on a beat-to-beat basis. Special all-purpose (SAP) and special high-sensitivity (SHS) collimator sections can be interchanged without movement of the patient. Thus, left-ventricular regions of interest delineated on SAP multigated images can be transferred to SHS dynamic images to generate beat-to-beat volume curves. Preliminary balloon studies demonstrated an excellent correlation between ejection fractions calculated with the two collimators: r > 0.99, n = 17, p < 0.001. Varying the volume of an adjacent "right ventricle" balloon failed to alter significantly the count rate from the "left ventricle" balloon's region on interest. Preliminary results on 12 patients, comparing standard-camera ejection fractions with average beat-to-beat ejection fractions, showed that it is possible to measure cardiac function on a beat-to-beat basis with a single-crystal gamma camera. There was minimal difference between the ejection fractions calculated by the dual-collimator method and a standard gated technique (r = 0.98, n = 12, p < 0.001).
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |