Unexpected Deadtime Losses in a Modified Rectilinear Scanning System

Philip H. Cooper, Samuel R. Lerner and Felix J. Pircher

VA Hospital, Houston, Texas

Correspondence: For reprints contact: Philip H. Cooper, V.A. Hospital, 2002 Holcomb Blvd., Houston, Tex. 77031.

ABSTRACT

Larger than expected deadtime losses were experienced with a rectilinear scanner interfaced with a multichannel analyzer in part due to inability to use the built-in livetime clock. Correction factors developed experimentally for these losses are presented.







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY THE JOURNAL OF NUCLEAR MEDICINE
Copyright © 1973 by the Society of Nuclear Medicine.